ISO 16063-21 Calibration & Traceability Review

Define selected-sensor calibration records, frequency points, uncertainty and laboratory scope. Request genuine model evidence and download a blank worksheet.

Request a matched configuration →

Comparison calibration and traceability are different records

ISO 16063-21 describes vibration calibration by comparison to a reference transducer. It provides a method framework; citing it does not prove that a particular sensor has been calibrated or that the issuing laboratory is accredited.

Traceability requires a documented chain to reference standards with relevant uncertainty. Ask for the actual method, reference standard identifiers, valid reference calibration and the uncertainty applicable to the measurement points. A generic certificate of conformity is a different document.

ISO 16063-21 scope and edition ↗
Concept comparison calibration: reference and test transducer, simultaneous measurement, uncertainty and traceability records
Concept process diagram. It does not depict an owned laboratory or a measured IEP test.

What to check on a selected-sensor certificate

Record fieldWhat the buyer should verify
IdentityExact model, serial number and X/Y/Z axis; the delivered sensor must match the record.
Method and setupComparison method, mounting condition, reference transducer and acquisition chain. Do not infer a laser or air-bearing system from a standards citation.
Sensitivity and unitsReference frequency, measured sensitivity, mV/g and/or mV/(m/s²), force or acoustic units where applicable.
Electrical conditionsExcitation current, relevant supply/compliance, bias and loading where these affect the result.
Frequency responseActual measured points and tolerance. Identify whether a curve is measured, nominal or a requested target.
UncertaintyExpanded uncertainty by point/band, coverage factor and stated conditions; no universal uncertainty value is assumed.
Conditions and validityCalibration date, laboratory environment, procedure and any recommended recalibration interval.
Traceability / accreditationIssuing laboratory identity, reference chain and, when required, current accreditation scope covering the method/range.

Request the level of documentation your procedure needs

Configuration and release data

Request the datasheet, dimensional drawing, connector/axis map and supported operating window for the selected identifier. These do not replace a calibration record.

Calibration and traceability

Request serial/axis-linked measured data, method, frequency points, reference chain and uncertainty. Availability and coverage need confirmation in the quotation.

Accredited calibration

If required by your quality procedure, request the issuing laboratory and applicable ISO/IEC 17025 scope. We do not claim site-owned accreditation or an available accredited upgrade without the specific laboratory arrangement.

A159.2Hz field check is not a full frequency calibration

A mechanical source such as the IEP-CAL-159 request can help check a sensor/cable/input chain at one defined frequency and load. The source needs its own amplitude, load and traceability evidence. A successful field reading does not establish broadband response, transverse sensitivity or calibration at extreme temperature.

Microphones need a suitable acoustic calibration procedure and source. An acceleration field source cannot serve as a microphone calibrator.

Review the mechanical field-source request →

TEDS readback should agree with the delivered record

Check model, serial number, calibrated sensitivity, units, calibration date and axis mapping against the actual record. TEDS storage does not perform calibration and does not establish traceability by itself. NI distinguishes Template 25 for IEPE acceleration/force from Template 27 for microphones with integrated preamplifiers.

Request a real readback from the selected sensor and intended host software when automatic identification is important. No illustrative screenshot is presented here as a verified IEP result.

NI TEDS template overview ↗

Engineering files: check the exact model release

Use the blank worksheet now. Request the selected configuration’s actual documents with the quote; no generic drawing or example certificate is presented as a released model record.

Selected-model release required

  • Datasheet and usable-response curve
  • Released STEP and dimensional / mounting drawing
  • Mating-view pinout and axis drawing
  • Calibration certificate and applicable scope
  • TEDS readback and environmental evidence
Review calibration and traceability requirements →

Supply your audit requirements with the inquiry

State the governing test procedure, required frequencies and uncertainty, acceptance rule, requested laboratory accreditation and document format. For triaxial sensors, clarify whether each axis needs separate calibrated values and uncertainty records. For temperature or pressure testing, request evidence under the relevant condition rather than assuming an ambient record covers it.

Is an example certificate available now?

No released IEP sample certificate is currently published. We provide a blank request worksheet and describe the record fields. Ask for a genuine, permitted selected-model sample or issuing-laboratory scope with your quotation.

Does the site claim an ISO 17025 laboratory?

No. Accredited calibration is a documentation request requiring confirmation of the issuing laboratory and applicable scope. Accreditation applies to the named organization and scope; it cannot be inferred from a sensor brand.

MODEL + INTERFACES + MATCHED ACCESSORIES

Request your complete measurement chain

Choose a configuration, enter actual unit quantities and request the documents needed for release. Specifications, accessory fit, price and delivery are confirmed in the quotation.

[email protected]

Instrument channel capacity or custom input requirements will be reviewed separately.

Matched accessory requests

Quantity is per line. Conditioner and calibration upgrades are optional. Enter your own planning allowance only if useful; blank amounts remain unpriced.

Select a configuration to see interface-specific accessory requests.

Engineering documents to request

Selected-model release files and certificate availability must be confirmed. This checklist does not imply that they already exist.

Optional planning allowances & budget

Known allowances: USD 0.00 · 1 unpriced lines.

Buyer-entered planning figures, not seller pricing. Freight, tax and other unpriced work are excluded.

Open prefilled email →

Review the complete request brief
IEPE Dynamic Instruments — configuration inquiry
Name: 
Company: 
Reply email: 
Requested model: Custom / to confirm
Configuration: Custom measurement chain
Requested sensitivity: To confirm
Connector request: To confirm
Pin assignment request: To confirm
Host DAQ / input: Not supplied
Sensor / instrument quantity: 1 units
Input channels required: confirm the complete instrument configuration
Original model: Not supplied
Submission page: /
Originating page: /
Model evidence: Custom requirements require review.
Accessory request lines:
Document requests: Selected-model datasheet; Calibration method, traceability and uncertainty
Buyer model allowance each: not supplied
Buyer target total budget: not supplied
Known planning allowances: USD 0.00; 1 unpriced lines. Buyer-entered estimates only; not a quotation or savings guarantee. Freight/tax/other unpriced work excluded.
Project requirements:
Please confirm selected-model specifications, compatible interfaces, document availability, formal pricing, payment terms and delivery in the quotation.